CYCLOTRON INSTITUTE logo
 [E-mail] [BibTeX] [RIS] [Request]
The dependence of the escape of characteristic X-rays from planar germanium detectors on the direction of the incident radiation
Pub type: Article
Citation: ILAKOVAC1984210
Status: Published
Journal: Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment
Volume: 228
Number: 1
Year: 1984
Month: December
Pages: 210 - 214
ISSN: 0168-9002
URL: http://www.sciencedirect.com/s...
DOI: 10.1016/0168-9002(84)90034-2
Abstract: Abstract The escape of germanium Kα X-rays from two planar germanium detectors for photons of 22.1 keV, incident at angles 0°, ± 10°, ṡ, ± 40°, was measured. Applications of the measurements are discussed.
Month_val: 12
Sig contrib: V. Horvat
Authors: Ilakovac, K.
Horvat, V.
Ilakovac, N.
Added by: [DGM]